Page 75
Choosing option 1 from the Calibration Selection Screen will display the
AUTO
CAL
screen shown below. The displayed mode may vary depending on the
instrument version and/or previous selections and also for GA2005H and
GA2005S Excieo models, MODE selection using
Soft Key 4
is not possible.
For the multi-purpose Excieo variation, use
Soft Key 4
labelled
MODE
to display
a popup which allows the selection of the available modes. Press
Soft Key 4
labelled
CANCEL
to return to the previous screen without changing the
calibration mode otherwise press
1
,
2
,
3
or
4
depending on which mode /
transducer you wish to calibrate.
To reset all displayed axes measurements press
Soft Key 3
labelled
RESET
.
The ‘
Cal To
’ figure displayed can be changed by selecting
Soft Key 2
labelled
CAL TO
. If selected then the ‘
Cal To
’ figure will be blanked and you will be
required to enter a new value using the numeric keypad. The ‘
Cal To
’ figure is
the value outputted from your calibrator in m/s
-2
. This figure can be found on
the calibration certificate of your calibrator.
Figure 87 – Automatic Calibration (Type H)
Figure 88 – Automatic Calibration Mode Selection
Figure 86 – Automatic Calibration (Type M)
09:44:02
26:08:11
HARM
AUTO CAL
RESET
CAL TO
09:44:00
26:08:11
HARM
AUTO CAL
RESET
EXIT CAL
TO
MODE
EXIT
Press number of axis
Press number of axis
Range: 0.02 – 100 ms
-2
3. Z:
0.130
1. X:
9.986
2. Y:
0.130
Cal To: 10.00 ms
-2
Range: 0.02 – 100 ms
-2
3. Z:
0.130
1. X:
9.986
2. Y:
0.130
Cal To: 10.00 ms
-2
09:44:05
26:08:11
HARM
AUTO CAL
CANCEL
Press number of axis
Range: 0.02 – 100 ms
-2
3. Z:
0.130
1. X:
9.986
2. Y:
0.130
Cal To: 10.00 ms
-2
1. HARM
2. WBV
3. Single
4. Custom