2. Installation Manual
- 61 -
•
Inserting annotations in the image can lead to problems
at the hospital. In the QCW, use annotations that are
outside of the image.
•
The designated film size is 11 x 14 inch film, and
automatic selection of the magazine and printing has
been confirmed.
•
Annotations in the image are problematic in the US and
EU. Thus, although DICOM annotation was used, it
was not printed. Although annotation can be transferred
without any errors in DICOM, an annotation error
occurred in the log when transferring from the gateway
to the 2180, and printing was not performed. Data was
transferred from KCR to the validation tool, and the
DICOM transfer method was compared to CXDI.
However, the only differences were in the image size,
aspect ratio, and annotation position.
•
Although the annotation position is 1 in terms of
conformance, it is 0 as sent from KCR. The result did
not change even after the change.
Agfa DryStar
3000
80
µ
m x 4256 x 5174
•
In the standard Agfa system, the annotation option is
selected. Therefore, the full image area will not be
printed correctly unless you ask the serviceman to
deselect the annotation option.
2128 x 2587 (@160
µ
m)
Imation DryView
8700+8800
78
µ
m x 4096 x 5220
Note: The 8700 printer cannot display in landscape
orientation. Also, the maximum density is 3.1. If the 8800
box is not added, this printer cannot be used by the
CXDI. However, it can be used starting from CXDI
version 2.0.
•
If there is an Imager Pixel Spacing tag (0018, 2264), the
imager will fail.
1996 x 2544 (@160
µ
m)
Fuji CR-DPL
/FM-DPL + FM-
PS551
100
µ
m x 3520 x 4280 (value after passing through the
print spooler)
•
The above settings are the allowable area size in a
configuration not using annotation.
•
Annotation will be supported from the next version.
Annotation in currently possible in US-ASCII only. IDs
are designated 1 to 6 and correspond to the top left, top
center, top right, bottom left, bottom center, and bottom
right, respectively. The maximum area size with
annotation support is 3500 x 4170 for 35 cm x 43 cm
and 2538 x 3522 for B4.
•
Use the Disable function for N-EVENT-REPORT to
disable this setting.
•
A function is provided for disabling the returning of
warning messages.
0107 (Attribute list error)
Return/Not return
0116 (Attribute Value out of Range)
Return/Not return
B604 (Image has been demagnified)
Return/Not return
•
The Film Size parameter can be used. The Media Type
(BLUE, CLEAR) parameter can also be used.
•
The Film Orientation parameter is fully supported.
•
The following presets are made so that LUT has the
2200 x 2675 (@160
µ
m)
Содержание CXDI-50C
Страница 15: ......
Страница 23: ...2 Installation Manual 6 3 Power Box assemble package 4 Grid optional ...
Страница 35: ...2 Installation Manual 18 When normal imaging When timeout due to RX_REQ not negating ...
Страница 63: ...2 Installation Manual 46 Fig 3 Input parameters ...
Страница 115: ...2 Installation Manual 98 2 Power Box Unit mm Mass 4 2Kg ...
Страница 116: ...2 Installation Manual 99 3 Remote Switch Unit mm Mass 0 5Kg 4 Grid Optional 508 27 452 Unit mm Mass 0 9Kg Frame ...
Страница 118: ......
Страница 138: ...4 Repair Guide 3 Do not touch the tabs when replacing the sensor cable Tabs in Enlarged View Sensor Cable ...
Страница 163: ......
Страница 167: ...CXDI SYSTEM 1 2 Rev 04 Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Canon Inc ...
Страница 186: ......
Страница 188: ......
Страница 189: ...Content 1 Gathering Information about Problems 1 2 How to back up and recover the system 3 ...