2. Installation Manual
- 60 -
Printer Model Specifications (Reference)
Name Specifications
Maximum equivalent
area in CXDI
Kodak MLP190
80
µ
m x 4096 x 5120
2048 x 2560 (@160
µ
m)
Kodak
KE
Kodak Print
Spooler Model
100
79
µ
m x 4090 x 5120 (value after passing through the
print spooler)
•
The above settings are the size of the effective area
when the image passes through the print spooler and
the image is plotted up to the annotation area. In other
words, these settings do not display an annotation area,
instead handling it as an image area.
•
If the data is transferred without setting [Image creation
in CXDI], the Requested Image Size setting is used in
DICOM. In this case, the annotation area is
automatically displayed in the 2180 printer. As a result,
a maximum image area of 79
µ
m x 4090 x 4996 must
be designated. In this case, the CXDI relies on the 2180
for image rotation (Film Orientation), but images larger
than 1.7 MB cannot be rotated by the 2180. Therefore,
operation without the setting for [Image creation in
CXDI] cannot be performed in the 2180. (Although
operation is possible by setting Requested Image Size
only for using DICOM without the setting for [Image
creation in CXDI], this option is not installed in the
CXDI.)
•
Although the resolution of the printer itself is 79
µ
m x
4090 x 5260, this complete resolution cannot be used
when the image passes through the spooler.
When “_” is used in AE Title, the association is
rejected.
Use the Disable function for N-EVENT-REPORT to
disable this setting.
The FilmSize parameter can be used. The MediaType
(BLUE, CLEAR) parameter is also supported.
•
Although the Film Orientation parameter is supported
up to 1.7 MB, in actuality, DR images cannot be
rotated. Like the DryView8700, the images must be
rotated by the CXDI side.
•
When Requested Image Size is expanded, the
maximum plotting size is limited (79 um x 4090 x
4996) so that the annotation area can be obtained.
When a Requested Image Size expansion error occurs,
the image is interpolated and printed at the suitable
size. In this case, the error does not return to the CXDI
side. For example, the image is printed at 310 mm even
if 326 mm is designated.
•
If an expansion error occurs in the spooler when Multi
Display Format is used, the print queue cannot be
processed.
•
The system is in a critical state when a Failure status is
indicated. A user message is displayed indicating this
state, and images are no longer transferred. (Fully
installed)
•
During the Warning status, image transfer is performed
while the user message is displayed (Fully installed).
•
If operation is aborted due to an error, a new
association could not be established when the data was
resent from the CXDI. GW was reset to recover the
error.
2018 x 2528(@160
µ
m)
Содержание CXDI-50C
Страница 15: ......
Страница 23: ...2 Installation Manual 6 3 Power Box assemble package 4 Grid optional ...
Страница 35: ...2 Installation Manual 18 When normal imaging When timeout due to RX_REQ not negating ...
Страница 63: ...2 Installation Manual 46 Fig 3 Input parameters ...
Страница 115: ...2 Installation Manual 98 2 Power Box Unit mm Mass 4 2Kg ...
Страница 116: ...2 Installation Manual 99 3 Remote Switch Unit mm Mass 0 5Kg 4 Grid Optional 508 27 452 Unit mm Mass 0 9Kg Frame ...
Страница 118: ......
Страница 138: ...4 Repair Guide 3 Do not touch the tabs when replacing the sensor cable Tabs in Enlarged View Sensor Cable ...
Страница 163: ......
Страница 167: ...CXDI SYSTEM 1 2 Rev 04 Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Canon Inc ...
Страница 186: ......
Страница 188: ......
Страница 189: ...Content 1 Gathering Information about Problems 1 2 How to back up and recover the system 3 ...