2. Installation Manual
- 91 -
(4) Turn on the irradiation ramp of the X-ray generator, cross line of the X-ray tube is cast over the
sensor unit. Align the position to meet the requirements described in 1) and 2) below.
1) Align the position so that the cross lines of X-ray tube and the sensor unit exactly match. If the
position is mismatched as shown in [NG] below, the sensor unit and the stand are not aligned
horizontally to the X-ray tube. In this case, put the spacer between the base of the stand and the
floor to align the slight differences.
[Top View]
[OK]
[NG]