2. Installation Manual
- 38 -
5-6) “Taskbar and Start Menu Properties” appears. Click “Start Menu” tab, and then click
Taskbar and Start Menu Properties
⇒
Start Menu
⇒
Classic Start
⇒
Menu Customize
5-7) The “Remove Shortcuts/Folders” dialog box appears after click the “Remove” button.
And double-click the “Startup folder”
5-8) Remove the “ccrstart.bat” item from it.
5-9) After “Remove” button clicked, the confirmation of deleting file appears. If you are
going to remove it, click “Yes” button.
5-10) After confirm that the “ccrstart.bat” item is removed from “Startup Group”, close all
the application on the desktop and re-login to Windows XP.
5-11) Make sure that the CXDI application software will not start automatically after login
to Windows XP. And then shutdown Windows XP, turn the CXDI system power off.
* When the CXDI application is deleted from the Start menu due to repair or other
reasons, be sure to always perform the procedures outlined in “Adding CXDI
application software onto the Start menu” and “Changing the window size” when
the repair is complete.
Содержание CXDI-50C
Страница 15: ......
Страница 23: ...2 Installation Manual 6 3 Power Box assemble package 4 Grid optional ...
Страница 35: ...2 Installation Manual 18 When normal imaging When timeout due to RX_REQ not negating ...
Страница 63: ...2 Installation Manual 46 Fig 3 Input parameters ...
Страница 115: ...2 Installation Manual 98 2 Power Box Unit mm Mass 4 2Kg ...
Страница 116: ...2 Installation Manual 99 3 Remote Switch Unit mm Mass 0 5Kg 4 Grid Optional 508 27 452 Unit mm Mass 0 9Kg Frame ...
Страница 118: ......
Страница 138: ...4 Repair Guide 3 Do not touch the tabs when replacing the sensor cable Tabs in Enlarged View Sensor Cable ...