2. Installation Manual
- 32 -
(2) Checking the Firmware Version
1) Purpose
1-1) Failing to use the proper versions of the firmware and PLD code with the CXDI
application can result in an error, and system operation cannot be guaranteed.
Therefore, the versions of the firmware must be checked to ensure that they are
correct. (The combination of this firmware refers to CXDI Software Combination
List.)
2) Notes
2-1) This check should always be performed at installation, and if necessary, the firmware
versions should be upgraded.
2-2) This check cannot be performed with only the control PC. Connect the imaging units
and other equipment, and start up in the normal imaging status.
3) Procedure
3-1) Checking the firmware alone
a. Start up the CXDI system.
b. Display the version information from the user mode.
SYSTEM > SETUP MENU > VERSION INFORMATION
c. Confirm the firmware version.
A. Hardware version
Product type and Sub No.: Product type is identified from Sensor serial No. is set to the
Imaging unit (07=50G and 0a=50C). Sub No. indicates the states of the dipswitch2-1 to 4.
B. Firmware initialization code version
This is the version of the initialization code written on the PWB-50Di. Initialization code
will be downloaded and settings will be reset to the default (factory) settings by turning
ON the power while pressing the initialization switch on the power box.
Display “1.1.0” on the screen indicates version 1.01.00.
C. Firmware normal code version
This is the version of the normal code installed on the PWB-50Di. Usually the system
operates with this code. It must be updated as required.
Display “1.1.00” on the screen indicates version 1.01.00.
A B
C
Содержание CXDI-50C
Страница 15: ......
Страница 23: ...2 Installation Manual 6 3 Power Box assemble package 4 Grid optional ...
Страница 35: ...2 Installation Manual 18 When normal imaging When timeout due to RX_REQ not negating ...
Страница 63: ...2 Installation Manual 46 Fig 3 Input parameters ...
Страница 115: ...2 Installation Manual 98 2 Power Box Unit mm Mass 4 2Kg ...
Страница 116: ...2 Installation Manual 99 3 Remote Switch Unit mm Mass 0 5Kg 4 Grid Optional 508 27 452 Unit mm Mass 0 9Kg Frame ...
Страница 118: ......
Страница 138: ...4 Repair Guide 3 Do not touch the tabs when replacing the sensor cable Tabs in Enlarged View Sensor Cable ...
Страница 163: ......
Страница 167: ...CXDI SYSTEM 1 2 Rev 04 Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Canon Inc ...
Страница 186: ......
Страница 188: ......
Страница 189: ...Content 1 Gathering Information about Problems 1 2 How to back up and recover the system 3 ...