2. Installation Manual
- 57 -
Examples of parameters used with different makers and types of printers (reference)
Name Default
Parameters
Significance
Kodak MLP190
-A 320 -T NO -M CUBIC –m
NORMAL -S CS000
(entered in param member)
80 (entered in pixelPitch
member)
14 x 17
4096 (entered in W member)
5120 (entered in H member)
•
Trimming OFF
•
Cubic spline interpolation
•
Smoothing: normal
•
Maximum density: 3.20
•
Curve shape 0 (density linear) as Config
Info
Kodak KE
Kodak Print Spooler
Model 100
-A 320 -T NO -M CUBIC –m
NORMAL -S CS000
(entered in param member)
79 (entered in pixelPitch
member)
14 x 17
4090 (entered in W member)
5120 (entered in H member)
11 x 14
3194 (entered in W member)
4096 (entered in H member)
•
Trimming OFF
•
Cubic spline interpolation
•
Smoothing: normal
•
Maximum density: 3.20
•
Curve shape 0 (density linear) as Config
Info
Agfa DryStar 3000
-A 320 -T NO -M CUBIC -m
140 -S
“PERCEPTION_LUT=200”
(entered in param member)
80 (entered in pixelPitch
member)
14x17
4256 (entered in W member)
5174 (entered in H member)
•
Trimming OFF
•
Cubic spline interpolation
•
Smoothing: slightly sharp (edges
emphasized)
•
Maximum density: 3.20
•
S
“PERCEPTION_LUT=200(LINEAR)”
(If the output fails to be linear with
“LINEAR”, on-site adjustments with the
printer manufacturer must be performed.
Kodak Imation DryView
8700 + Pacs LINK IMN
9410
-A 310 -T NO -M CUBIC -
m ? -S LUT=0?, 2? (?: Site
Dependent)
(entered in param member)
78 (entered in pixelPitch
member)
14x17
4096 (entered in W member)
5220 (entered in H member)
•
Trimming OFF
•
Cubic spline interpolation
•
Smoothing must be adjusted at the user’s
site.
•
Maximum density: 3.10
•
S LUT = m, n is designated as the
Config Info but m and n are adjusted on-
site by the Kodak service engineer.
Basically, adjustment is performed to
achieve a linear output.
Kodak Imation DryView
8700 + GW
-A 310 -T NO -M CUBIC -
m ? -S LUT=?, ? (?: Site
Dependent)
(entered in param member)
78 (entered in pixelPitch
member)
14x17
4096 (entered in W member)
5220 (entered in H member)
•
Trimming OFF
•
Cubic spline interpolation
•
Smoothing must be adjusted at the user’s
site.
•
Maximum density: 3.10
•
S LUT = m, n is designated as the
Config Info but m and n are adjusted on-
site by the Kodak service engineer.
Basically, adjustment is performed to
achieve a linear output.
Kodak Imation DryView
8700+8800
-A 320 -T NO -M CUBIC –S
“LUT=m, n”-m (on-site
adjustment)
(entered in param member)
78 (entered in pixelPitch
member)
4096 (entered in W member)
5220 (entered in H member)
•
Trimming OFF
•
Cubic spline interpolation
•
Maximum density: 3.20
•
S LUT = m, n is designated as the
Config Info but m and n are adjusted on-
site by the Kodak service engineer.
Basically, adjustment is performed to
achieve a linear output.
Содержание CXDI-50C
Страница 15: ......
Страница 23: ...2 Installation Manual 6 3 Power Box assemble package 4 Grid optional ...
Страница 35: ...2 Installation Manual 18 When normal imaging When timeout due to RX_REQ not negating ...
Страница 63: ...2 Installation Manual 46 Fig 3 Input parameters ...
Страница 115: ...2 Installation Manual 98 2 Power Box Unit mm Mass 4 2Kg ...
Страница 116: ...2 Installation Manual 99 3 Remote Switch Unit mm Mass 0 5Kg 4 Grid Optional 508 27 452 Unit mm Mass 0 9Kg Frame ...
Страница 118: ......
Страница 138: ...4 Repair Guide 3 Do not touch the tabs when replacing the sensor cable Tabs in Enlarged View Sensor Cable ...
Страница 163: ......
Страница 167: ...CXDI SYSTEM 1 2 Rev 04 Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Canon Inc ...
Страница 186: ......
Страница 188: ......
Страница 189: ...Content 1 Gathering Information about Problems 1 2 How to back up and recover the system 3 ...