
•
NO
(default) sets the Cadex analyzer not to ask for the battery
C-code.
6.
Press
ENTER
.
7.
Press
or
to scroll to accept changes and press
ENTER
.
8.
Press
ESC
twice to return to the global display.
Performing an Automatic OhmTest
Every time the Cadex analyzer runs a charge cycle, it can automatically perform an
OhmTest on the battery as well (except when the charge cycle is being performed
as part of a custom program). This test can be turned off.
Note
Automatic OhmTest cannot be turned on or off using the following
programs: QuickTest, Q-Learn, Learn and Life-Cycle. OhmTest is included
in these programs and the settings cannot be changed.
To turn Automatic OhmTest on or off
1.
From the global display, press
MENU
.
2.
Press
or
to scroll to
PROGRAM SETTINGS
and then press
ENTER
.
3.
Press
or
to scroll to
AUTOMATIC
O
HM
T
EST
and then press
ENTER
.
4.
Press
or
to scroll to the required setting:
•
YES
(default) sets the OhmTest to run automatically before every
charge cycle (except during custom programs).
•
NO
turns off the automatic OhmTest.
5.
Press
ENTER
.
6.
Press
or
to scroll to
YES
to accept changes and press
ENTER
.
7.
Press
ESC
twice to return to the global display.
Cadex Electronics Inc. C7000-C Series Battery Analyzers
.
78