Parameter Setting
Max. Standby Voltage 2.35V/cell
Max. Charge Voltage 2.60V/cell
End-Of-Discharge 1.62V/cell
Gell SLA Batteries
Use the SLA default settings for Gell batteries.
Managing C-codes
If the
ASK FOR BATTERY
C-
CODE
option
is set to
YES
, the desired C-code must be
selected and the appropriate parameters entered before a battery is inserted or
before battery processing begins (see
To set the user input required by the
Cadex analyzer before battery service begins
Select a C-code
1.
From the global display, press the station key.
2.
Press
or
to scroll to the C-code to select.
3.
Press
ENTER
.
4.
Press
or
to scroll to
YES
to accept changes and press
ENTER
.
5.
Press
ESC
twice to return to the global display.
The “*” appears beside the C-code number on the display to indicate that the C-
code is active.
Display a C-code
1.
From the global display, press the station key.
2.
Use the
or
key to scroll to the C-code to display. Basic C-code
parameters appear on the same screen.
61