4.
Remove and examine the fuse carefully.
If the filament inside the glass tube is broken, replace the fuse. Be aware that the
filament can sometimes appear intact even when it is broken.
5.
Close the fuse cover by pushing and turning it clockwise with a small
coin or screwdriver until it locks in place.
6.
Insert the power cord and turn on the analyzer.
Upgrading the Firmware
From time to time, Cadex offers firmware upgrades for the analyzer. These can
usually be downloaded to from www.cadex.com and then copied to the analyzer
using the Cadex Firmware Utility software, available from Cadex Electronics Inc. (the
Firmware Utility software comes with BatteryShop
)
.
The current version of the firmware is always displayed in the lower left-hand
corner of the analyzer startup screen, followed by the current version of the Cadex
Firmware Utility software, as shown here:
C7400-C (C)2006 CADEX ELECTRONICS INC.
V1.00/1.00 ** DIAGNOSTICS PASSED **
Firmware Version
To verify firmware version (display start-up screen)
1.
From the global display, press
FN
and then press
7
.
2.
Press
ESC
to return to the global display.
Resetting the System
To reset parameters to the factory default and restart the analyzer
1.
Remove all batteries and adapters.
2.
Power-off the analyzer.
3.
Hold down the
MENU
key.