ADAPTER HANDLING PRECAUTIONS
•
Do not touch the adapter’s gold contacts at the rear or on
the analyzer station adapter connector. The static charge
can damage microcircuits on the adapter.
•
Do not remove the adapter from the analyzer while it is
servicing a battery. Remove the battery first and then the
adapter.
•
ESD HANDLING PRECAUTIONS: When the adapter is NOT
installed in analyzer:
•
The battery should not be connected to the adapter
•
Do not allow anything to touch the gold fingers on the
adapter connector card
•
Do not allow anything to touch the adapter connector in
the analyzer's adapter well
Each adapter contains up to ten battery service parameters called Configuration
Codes (or
C-codes
) in non-volatile memory. C-codes will not be lost when adapters
are removed and reinserted into any station on the analyzer or when the analyzer
is turned off.
Custom adapters are pre-programmed with C-codes for the most common
batteries. If a battery C-code is not available, program a new C-code or reprogram
an existing C-code into the analyzer.
Universal adapters, designed for generic batteries, are programmed with blank C-
codes and will display N
ULL
C
ODE
,
page 62). Battery Adapters can be installed, removed or reprogrammed while other
stations are being used. Once the adapter is inserted, the battery station receiving
the adapter automatically uses the last selected C-code.
Cadex Electronics Inc. C7000-C Series Battery Analyzers
.