Detailed
Message
Code Reasons
Solutions
BAD FUSE OR
DRIVER
160
The analyzer is not able to
supply current to the battery
through its current drivers.
Very low values of current
(about 0 mA) are displayed
on the analyzer and after 10
seconds processing stops.
Clean the battery and adapter
contacts.
Verify that mAh rating of the
battery matches the C-code.
Lower the Charge rate and
the Trickle Charge rate in the
C-code.
Use I/S settings (see
Intrinsically Safe Batteries
,
page 60) if the battery is
intrinsically safe.
The battery may be faulty. Try
a different battery in the same
station.
Single Cell NiCd or NiMH
batteries may produce this
error on the C7400ER-C.
Service single cells on the
C7200-C or C7400-C.
Check the fuses on the
analyzer. If all fails, contact
Cadex for service.
BATSHOP MODE
WAIT
34
Battery is detected on an
analyzer connected to
Cadex BatteryShop™.
Waiting for user to start
battery service in Cadex
BatteryShop.
If service has been started
and this message remains,
verify the C-code in Cadex
BatteryShop to ensure that
settings are correct for the
battery.
BATTERY
INSERTED
20
The Cadex 7200 detects a
battery in an adapter.
If message persists, press
ESC.