Program Parameters
There are a number of settings available that affect the way the Cadex analyzer
functions in most or all of its programs.
Starting Battery Service
By default, when a battery is inserted into the Cadex analyzer, the analyzer
prompts to verify the C-code before service starts. This verification can be turned
off so users can run a battery service as soon as a battery is inserted as follows:
•
The Cadex analyzer asks for Battery ID number and asks to verify C-code
settings.
•
The Cadex analyzer asks for Battery ID number only.
•
The Cadex analyzer asks to verify C-code settings only (default).
•
The Cadex analyzer starts the program using the active C-code without asking to
enter the Battery ID or verify the C-code.
To verify C-code or enter Battery ID before battery service begins
1.
From the global display, press
MENU
.
2.
Press
or
to scroll to
OPTION CONTROLS
and then press
ENTER
.
3.
Press
or
to scroll to
ASK FOR BATTERY ID
#
4.
Press
or
to scroll to the required setting:
•
YES
sets the Cadex analyzer to ask for the Battery ID number.
•
NO
(default) sets the Cadex analyzer not to ask for Battery ID
number.
5. Press
or
to scroll to
ASK FOR BATTERY
C-
CODE
5.
Press
or
to scroll to the required setting:
•
YES
sets the Cadex analyzer to ask the user to verify C-code
settings.
77