ABORt (short form -> AB
)
DESCRIPTION:
This stops a measurement.
SYNTAX:
ABORt
PARAMETERS:
None
QUERY FORM:
None
RESPONSE:
—
REMARKS:
This command also exists in a special short form:
AB
RESTRICTION:
This command is not permissible in the calibration mode.
6.3.6 Measurement Instructions
FETCh? (Kurzform -> FE?)
DESCRIPTION:
This requests a measured resistance value, which is subsequently
written to the output buffer.
SYNTAX:
FETCh?
PARAMETERS:
None
QUERY FORM:
Only query form
RESPONSE:
The measured resistance value with a unit. and valuation
For example: 134.75OHM
REMARKS:
This command also exists in a special short form:
FE?
RESTRICTION:
This command is not permissible in the calibration mode or if a
measurement has been stopped in the absence of any measured
values.
6.3.7 MEMory Subsystem
MEMory:STATe:DEFine
DESCRIPTION:
This assigns an ID to a memory number (device setting).
SYNTAX:
MEMory:STATe:DEFine <ID>,<number>
PARAMETERS:
Identification: 10-character ASCII string
Number: Memory number from 0 to 32
QUERY FORM:
MEMory:STATe:DEFine? <ID>
RESPONSE:
The memory number (0 to 32) related to the ID is
returned.
RESTRICTION:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.