SENSe:TCOMpensate:TEMPerature
DESCRIPTION:
This sets the temperature for manual temperature compensation.
In the query form you get the present temperature.
SYNTAX:
SENSe:TCOMpensate:TEMPerature <parameter>
PARAMETERS:
Temperature value, optionally with a unit (C or CEL)
QUERY FORM:
SENSe:TCOMpensate:TEMPerature?
RESPONSE:
The set temperature value and a corresponding unit (CEL) or the
topical temperature at autom. temp. measurement.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.
SENSe:FRESistance:LOAD
DESCRIPTION:
This selects the type of test object (R or Z).
SYNTAX:
SENSe:FRESistance:LOAD <parameter>
PARAMETERS:
REAL -> R is set (purely ohmic load)
COMPlex -> Z is set (inductive component)
QUERY FORM:
SENSe:FRESistance:LOAD?
RESPONSE:
REAL -> If R is set.
COMP -> If Z is set.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.
SENSe:AVERage:COUNt
DESCRIPTION:
This sets the number of averaging operations.
SYNTAX:
SENSe:AVERage:COUNt <parameter>
PARAMETERS:
Number of averaging operations. Value range: 1 to 100
QUERY FORM:
SENSe:AVERage:COUNt?
RESPONSE:
The set number of averaging operations.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.