101
of 148
CALCulate:LIMit:GW3
DESCRIPTION:
This specifies limiting value 3 (second largest value GW3)
when 4 comparator limits are selected.
SYNTAX:
CALCulate:LIMit:GW3<parameter>
PARAMETERS:
Limit 3 with an optional unit. Valid units: UOHM, MOHM, OHM, KOHM,
MAOHM. If no unit is specified, the unit of the value is interpreted as
OHM.
QUERY FORM:
CALCulate:LIMit:GW3?
RESPONSE:
Limit 3 of 4 selected comparator limits (value with unit).
BEMERKUNGEN:
The value transferred with this command needs to be acknowledged
with CALCulate:LIMit:ACKNowledge. The validity of this value is tested
with respect to the remaining limits.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.
CALCulate:LIMit:GW4
DESCRIPTION:
This specifies limiting value 4 (largest value GW4) when 4 comparator
limits are selected.
SYNTAX:
CALCulate:LIMit:GW4<parameter>
PARAMETERS:
Limit 4 with an optional unit. Valid units: UOHM, MOHM, OHM, KOHM,
MAOHM. If no unit is specified, the unit of the value is interpreted as
OHM.
QUERY FORM:
CALCulate:LIMit:GW4?
RESPONSE:
Limit 4 of 4 selected comparator limits (value with unit).
BEMERKUNGEN:
The value transferred with this command needs to be acknowledged
with CALCulate:LIMit:ACKNowledge (page 103). The validity of this value
is tested with respect to the remaining limits.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the
device is in the calibration mode.