114
of 148
ACCess:CONTrast
DESCRIPTION:
This enables/disables access to the contrast-setting menu.
SYNTAX:
ACCess:CONTrast <parameter>
PARAMETERS:
1 or ON -> Enables access to the contrast-setting menu.
0 or OFF -> Disables access to the contrast-setting menu.
QUERY FORM:
ACCess:CONTrast?
RESPONSE:
1 -> If access to the contrast-setting menu is enabled.
0 -> If access to the contrast-setting menu is disabled.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the device
is in the calibration mode.
ACCess:STATe
DESCRIPTION:
This enables/disables access to the status-display menu.
SYNTAX:
ACCess:STATe <parameter>
PARAMETERS:
1 or ON -> Enables access to the status-display menu.
0 or OFF -> Disables access to the status-display menu.
QUERY FORM:
ACCess:STATe?
RESPONSE:
1 -> If access to the status-display menu is enabled.
0 ->If access to the status-display menu is disabled.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the device
is in the calibration mode.
ACCess:REGister
DESCRIPTION:
This enables/disables access to the spare input/output menu.
SYNTAX:
ACCess:REGister <parameter>
PARAMETERS:
1 or ON -> Enables access to the spare input/output menu.
0 or OFF -> Disables access to the spare input/output menu.
QUERY FORM:
ACCess:REGister?
RESPONSE:
1 -> If access to the spare input/output menu is enabled.
0 -> If access to the spare input/output menu is disabled.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the device
is in the calibration mode.