118
of 148
ACCess:MANRange
DESCRIPTION:
This enables/disables access to manual range selection.
(From the main menu: RGE [
↑
] and RGE [
↓
] keys)
SYNTAX:
ACCess:MANRange <parameter>
PARAMETERS:
1 or ON -> Enables access to manual range selection.
0 or OFF -> Disables access to manual range selection.
QUERY FORM:
ACCess:MANRange?
RESPONSE:
1 -> If access to manual range selection is enabled.
0 -> If access to manual range selection is disabled.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the device
is in the calibration mode.
5.3.14 DATalogger Subsystem
DATalogger:STATe
DESCRIPTION:
This starts/stops the datalogger.
SYNTAX:
DATalogger:STATe <parameter>
PARAMETERS:
1 or ON -> The datalogger is started.
0 or OFF -> The datalogger is stopped.
QUERY FORM:
DATalogger:STATe?
RESPONSE:
1 -> If the datalogger is active.
0 -> If the datalogger is inactive.
RESTRICTIONS:
This command is not permissible if the device is in the calibration mode.
DATalogger:MINimum?
DESCRIPTION:
This calculates and fetches the minimum value of a datalogger block.
SYNTAX:
DATalogger:MINimum? <parameter>
PARAMETERS:
The datalogger block number (0 to 31).
QUERY FORM:
Only query form
RESPONSE:
The minimum value in MOHM, OHM or KOHM.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the device
is in the calibration mode.