60
of 148
Note:
The set output bits
cannot
be stored as part of the device setting.
[F3] key (HOME):
This effects a return to the main menu.
[F4] key (RETUrn):
This effects a return to the parameter selection menu.
5.4.4.11 Scaling menu
The various inputs for temperature measurement are scaled in this menu.
A Pt100 input for directly connecting a Pt100 sensor (e.g. type 2392-V001) is available in addition to a linear
voltage/current input for connecting a temperature transmitter (e.g. pyrometer).
The following entries are possible:
[F3] key (HOME):
This effects a return to the main menu.
[F4] key (RETUrn):
This effects a return to the parameter menu.
The [
↑
] and [
↓
] cursor keys are used to select the various inputs.
The selected parameter is displayed inversely.
The [POS1] key is used to select the 1st parameter in the window.
The [END] key is used to select the last parameter in the window.
Pressing the [F 1] key (SEL) invokes the menu for scaling the selected input.
Scaling Pt100 input
This menu is used to enter the coefficients for the Pt100 curve in the range
≥
100
Ω
(i.e. only positive
temperatures).
[F3] key (HOME): This effects a return to the main menu.
[F4] key (RETUrn): This effects a return to the scaling main menu.
Display of the scaling menu
SEL
HOME RETU
SCALED INPUTS
Pt100 INPUT
VOLTAGE INPUT
CURRENT INPUT
Display of the menu
for scaling the Pt100 input
INPUT
HOME RETU
Pt100 INPUT
Rt = Ro* (1+A*t+B*t*t)
Ro = 100
A = 0.0039083
B = -5.77e-07