27
of 148
5.3 Main menu
5.3.1 Measurement stopped
200 OHM 1mA 23.4C
112.45
Ω
0 5 SINGLE AUTO R
DATA COMP ENVE PARA
S 1.1 S 1.2 S 1.3
F 1 F 2 F 3 F 4
S 2.1 S 2.2 S 2.3 S 2.4 S 2.5 S 2.6
→
Status line 1
→
Status line 2
→
Meas. value
field
→
Footer
Display with a measurement stopped
Display in status line 1:
S 1.1 Measuring range
S 1.2 Measuring current or 20 mV with dry-circuit measurement
S 1.3 Temperature with temperature compensation or
setpoint
value
with
∆
% display
Display in status line 2:
S 2.1 Sample character (a sequential number from 0 to 9,
only while a measurement is in progress)
S 2.2 Error status (see page 138) or comparator evaluation
S 2.3 The memory number loaded last
S
2.4
Continuous
or
single-frame
measurement
S 2.5 Manual or automatic range selection
S 2.6 R or Z measurement procedure
Display in the measurement-value field: The value measured last
Display in the footer:
Designation of the function keys
The following entries are possible in this menu:
[F1] key (DATA):
This invokes the datalogger evaluation and configuration menus.
[F2] key (COMP):
This invokes the comparator evaluation and configuration menus.
[F3] key (ENVE):
This invokes the max./min, evaluation menu.
[F4] key (PARA):
This invokes the parameter selection menu.
[ENT] key:
Pressing the ENTER key invokes the code input menu.
[START] key:
Pressing this key commences a measurement.
[A/M] key:
This key is used to switch over between automatic and manual range
selection.
The [RGE
↑
] and [RGE
↓
] keys are used to select ranges manually in the corresponding mode (not in the case
of dry-circuit measurements).