45
of 148
Selecting a
FILTER
:
The [
→
] and [
←
] cursor keys are used to set the following values:
ALL
:
Save all values.
GOOD
:
Only save good values (those within the comparator limits).
BAD
:
Only save bad values (those outside the comparator limits).
Xth VALUE
:
Save every Xth measured value.
hh:mm:ss
:
Save at the time interval specified in hours, minutes and seconds.
dR xx.xxx
Ω
:
Save a value if its difference to the previous value exceeds dR.
Setting the X parameter:
Pressing the [F1] key (INPUT) invokes the input mode.
Numeric keys [0] to [9] are used to enter the required value.
Value range: 2 to 9999
The [BSP] key is used to delete the character entered last.
The [C] key is used to delete all the entered characters and repeat the input procedure from the beginning,
without exiting the input mode.
The [ESC] key is used to cancel the current entry, and exit the input mode.
The [ENT] key is used to save the input value - provided that it is valid - and exit the input mode.
Setting the time interval:
Pressing the [F1] key (INPUT) invokes the input mode.
Numeric keys [0] to [9] are used to enter the required value.
Value range: 00:00:01 to 99:59:59
The [BSP] key is used to delete the character entered last.
The [C] key is used to delete all the entered characters and repeat the input procedure from the beginning,
without exiting the input mode.
The [ENT] key is used to save the input value - provided that it is valid - and exit the input mode.
The [ESC] key is used to cancel the current entry, and exit the input mode.
The [F1] key (s), [F2] key (m) or [F3] key (h) is used to save the input value - provided that it is valid - and exit
the input mode.
Specifying the dR parameter:
Pressing the [F1] key (INPUT) invokes the input mode.
The function keys are redefined in this case.
Numeric keys [0] to [9] and the [.] key are used to enter the dR value.
(5 digits plus the decimal point)
Value range: 0.01 m
Ω
to 200 k
Ω
The [BSP] key is used to delete the character entered last.
The [C] key is used to delete all the entered characters and repeat the input procedure from the beginning,
without exiting the input mode.
The [ESC] key is used to cancel the current entry, and exit the input mode.
The [F1] key (mOHM), [F2] key(OHM) or [F3] key(kOHM) is used to save the input value - provided that it is
valid - and exit the input mode.
mOHM OHM kOHM