ACCess:CALibration
DESCRIPTION:
This enables/disables access to the calibration menu.
SYNTAX:
ACCess:CALibration <parameter>
PARAMETERS:
1 or ON -> Enables access to the calibration menu.
0 or OFF -> Disables access to the calibration menu.
QUERY FORM:
ACCess:CALibration?
RESPONSE:
1 -> If access to the calibration menu is enabled.
0 -> If access to the calibration menu is disabled.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the device
is in the calibration mode.
ACCess:DATalogger
DESCRIPTION:
This enables/disables access to the datalogger menu.
SYNTAX:
ACCess:DATalogger <parameter>
PARAMETERS:
1 or ON -> Enables access to the datalogger menu.
0 or OFF -> Disables access to the datalogger menu.
QUERY FORM:
ACCess:DATalogger?
RESPONSE:
1 -> If access to the datalogger menu is enabled.
0 -> If access to the datalogger menu is disabled.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the device
is in the calibration mode.
ACCess:SCALe
DESCRIPTION:
This enables/disables access to the scaling menu.
SYNTAX:
ACCess:SCALe <parameter>
PARAMETERS:
1 or ON -> Enables access to the scaling menu.
0 or OFF -> Disables access to the scaling menu.
QUERY FORM:
ACCess:SCALer?
RESPONSE:
1 -> If access to the scaling menu is enabled.
0 -> If access to the scaling menu is disabled.
RESTRICTIONS:
This command is not permissible if a measurement is in progress or the device
is in the calibration mode.