![burmeier Regia Скачать руководство пользователя страница 97](http://html1.mh-extra.com/html/burmeier/regia/regia_instruction-manual_2832795097.webp)
Step
Buttons
Action
1
Press and hold both of the marked
keys simultaneously (a rapid inter-
mittent signal sounds) until after
approximately 5 seconds a slow
intermittent signal sounds (=RE-
SET/ manual mode)
Note: When the lift motors are at
their end position (top/bottom), the
initialisation is complete. Other-
wise, follow step 2 below.
2
Raise the mattress base height
with the “Height UP” button to the
highest horizontal position until
both motors switch off automati-
cally at the highest position, and
keep the button pressed for a fur-
ther 2 seconds. When the button
is released, the intermittent sound
signal stops.
End of initialisation.
If no adjustments are possible and intermittent sound signals sound instead, this means that
the control unit has detected a fault and is locked. In this case, reset the control unit (see
).
Replacement of electrical components
Part B: Operator and Technical Personnel
61
Содержание Regia
Страница 2: ...Part A General Information ...
Страница 3: ...Regia Standard care bed Regia care bed DSG optional ...
Страница 6: ...II ...
Страница 30: ...Part B Operator and Technical Per sonnel ...
Страница 31: ...Regia Standard care bed Regia care bed DSG optional ...
Страница 36: ...IV ...
Страница 105: ...Appendix Part B Operator and Technical Personnel 69 ...
Страница 106: ...Part C Care staff and residents ...
Страница 107: ...Regia Standard care bed Regia care bed DSG optional ...
Страница 151: ...Cleaning and Disinfection Part C Care staff and residents 41 ...