Ports for signal input/signal output parts
Phenomenon
EMC basic standard
Immunity level (test + compli-
ance)
Home healthcare environment
Electrostatic discharge (ESD)
IEC 61000-4-2
+/- 8 kV; contact
+/- 2 kV, +/- 4 kV¸ +/- 8 kV, +/- 15 kV;
+/- 25kV air;
Electrical fast transient distur-
bances/bursts
IEC 61000-4-4
+/- 1 kV; 100 kHz repetition frequency
Conducted interference in-
duced by high-frequency
fields
IEC 61000-4-6
3 V; 0.15 MHz to 80 MHz; 6V in ISM
and amateur radio frequency bands
between 0.15 MHz and 80MHz 80%
AM at 1 kHz
Table zz: Test specifications for the immunity of sheathings to high-frequency wire-
less communication equipment
Test fre-
quency
MHz
Frequen-
cy band
Radio serv-
ice
Modulation
Max.
power W
Distance
m
Immunity
test level
v/m
385
380 to 390
TETRA 400
Pulse modu-
lation 18 Hz
1.8
0.3
27
450
430 to 470
GMRS 460
FRS460
FM +/- 5%
deviation,
1kHz sine
wave
2
0.3
28
710
704 to 787
LTE band 13,
17
Pulse modu-
lation 217 Hz
0.2
0.3
28
745
780
810
800 to 960
GSM
800/900 TET-
RA 800
iDEN820,
CDMA 850,
LTE band 5
Pulse modu-
lation 18 Hz
0.2
0.3
28
870
930
1720
1700 to
1990
GSM 1800
CDMA 1900,
Pulse modu-
lation 18 Hz
2
0.3
28
Product Description
22
Part A: General Information
Содержание Regia
Страница 2: ...Part A General Information ...
Страница 3: ...Regia Standard care bed Regia care bed DSG optional ...
Страница 6: ...II ...
Страница 30: ...Part B Operator and Technical Per sonnel ...
Страница 31: ...Regia Standard care bed Regia care bed DSG optional ...
Страница 36: ...IV ...
Страница 105: ...Appendix Part B Operator and Technical Personnel 69 ...
Страница 106: ...Part C Care staff and residents ...
Страница 107: ...Regia Standard care bed Regia care bed DSG optional ...
Страница 151: ...Cleaning and Disinfection Part C Care staff and residents 41 ...