Problem
Possible causes
Solution
Continuous alarm
sounds during bat-
tery mode
•
Battery capacity deple-
ted
•
Connect bed to the mains supply to re-
charge the battery as soon as possible
Operation with suf-
ficiently charged
battery only possi-
ble for a short time
•
Limit of battery’s service
life reached
•
Replace the battery. Inform your operator if
repairs are necessary
Operation is not
possible despite
proper power sup-
ply
•
Control unit has shut
down due to overheat-
ing
•
The control unit has de-
tected a fault and for
safety reasons has
locked the adjusting
functions
•
Control unit defective
•
Handset faulty
•
Drive connecting cable
defective
•
Observe max. duty cycle: Intermittent duty
2 min ON/18 min OFF; replace the control
unit.
•
Perform RESET (see
•
Replace the control unit. Inform your oper-
ator if repairs are necessary
•
Replace the handset with a new one
•
Replace connecting cable
Manual CPR re-
lease of backrest
(optional accesso-
ry) is not possible
•
Bowden cable is too
loose or not attached
•
Bowden cable is kinked
•
Readjust at the release lever or attach
properly
•
Install new Bowden cable.
•
Inform your operator if repairs are neces-
sary
drive runs for a
brief time only,
then stops
•
Drive overloaded
•
Structural obstructions
in the way of bed adjust-
ment
•
Remove the overload in the bed, retest
•
Remove obstructions; move bed away
from obstructions (e.g. window sills, slant-
ing ceilings)
Control partly not
functioning
•
One or more motors are
not connected properly/
electrical plug connec-
tions have come loose
•
There is a serious prob-
lem with the control unit
For safety reasons, all
functions are locked
•
Check electrical connection of all motors/
plug connections;
•
Back rest, thigh rest and lift all function in-
dependently
Troubleshooting
Part C: Care staff and residents
33
Содержание Regia
Страница 2: ...Part A General Information ...
Страница 3: ...Regia Standard care bed Regia care bed DSG optional ...
Страница 6: ...II ...
Страница 30: ...Part B Operator and Technical Per sonnel ...
Страница 31: ...Regia Standard care bed Regia care bed DSG optional ...
Страница 36: ...IV ...
Страница 105: ...Appendix Part B Operator and Technical Personnel 69 ...
Страница 106: ...Part C Care staff and residents ...
Страница 107: ...Regia Standard care bed Regia care bed DSG optional ...
Страница 151: ...Cleaning and Disinfection Part C Care staff and residents 41 ...