![burmeier Regia Скачать руководство пользователя страница 93](http://html1.mh-extra.com/html/burmeier/regia/regia_instruction-manual_2832795093.webp)
WARNING
7.2
Replace the handset with a new one
•
The plugs for the components are connected to the appropriate control unit. To pre-
vent the plugs from being inadvertently disconnected, they are secured with a locking
device. This device can be carefully lifted off using a screwdriver if necessary.
•
The control unit sockets should be lightly greased inside with Vaseline. The plugs can
then be inserted more easily and the O-rings provide a better seal.
•
The locking device must always be properly refastened.
•
For information on the connection of various optional electrical components, such as
the reading lamp, please refer to the accessory instructions supplied.
In order to reconnect the electrical components correctly after replacement, it is essential
to observe the information in the chapter
Part A: Electrical connection diagram
.
Risk of injury
Failure to heed this warning may result in physical injury due to faulty maintenance.
•
All drive components are maintenance-free and must not be opened. In the event of a
malfunction, the corresponding components should always be replaced in full!
•
When replacing individual components, make sure that the plugs have undamaged O-
rings (for sealing) and are inserted into the control unit as far as they will go. This is the
only way to ensure proper sealing and faultless operation.
•
Ensure the polarity is correct when inserting plugs and do not apply excessive force if the
plug does not fit. The plugs have a groove and therefore only fit into the corresponding
connection sockets with one polarity.
•
Do not wrongly connect the motor connections on the control unit. This can lead to mal-
functions or even result in mechanical damage to the drives due to the system not switch-
ing off at the end position.
•
After replacing the control units and/or the attached drives for adjusting the mattress base
height, always initialise the control unit (= re-align the electronic path measurement for
these drives). This sets a new reference point in the control unit for correctly measuring
the path and avoids faults or damage to the lifting mechanism.
Replacement of electrical components
Part B: Operator and Technical Personnel
57
Содержание Regia
Страница 2: ...Part A General Information ...
Страница 3: ...Regia Standard care bed Regia care bed DSG optional ...
Страница 6: ...II ...
Страница 30: ...Part B Operator and Technical Per sonnel ...
Страница 31: ...Regia Standard care bed Regia care bed DSG optional ...
Страница 36: ...IV ...
Страница 105: ...Appendix Part B Operator and Technical Personnel 69 ...
Страница 106: ...Part C Care staff and residents ...
Страница 107: ...Regia Standard care bed Regia care bed DSG optional ...
Страница 151: ...Cleaning and Disinfection Part C Care staff and residents 41 ...