![burmeier Regia Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/burmeier/regia/regia_instruction-manual_2832795022.webp)
4.5.9
Electrical connection diagram
External switch mode power supply:
With this control unit, the safety extra-low voltage supply cable that leads to the external
switch mode power supply is permanently connected and cannot be disconnected. This
prevents a 230V mains cable from being inadvertently connected.
Motors for mattress base height
Duty cycle
Intermittent duty: 2 min ON / 18 min
OFF
Protection category
IP X4
Backrest motor
Type
Linak LA 27
Force/installation dimension/lift
4000 N / 438 mm / 200 mm
Input voltage
Max. DC 34 V
Duty cycle
Intermittent duty: 2 min ON / 18 min
OFF
Protection category
IP X4
Thigh rest motor
Type
Linak LA 27
Force/installation dimension/lift
4000 N / 272 mm / 70 mm
Input voltage
Max. DC 34 V
Duty cycle
Intermittent duty: 2 min ON / 18 min
OFF
Protection category
IP X4
All plugs are connected to the control unit. To prevent the plugs from being inadvertently dis-
connected, they are secured with a plug cover. Before replacing a plug, lift this cover careful-
ly using a screwdriver.
This bed can be delivered with standard equipment and also with optional equipment. In the
following chapters you will find the connection diagrams of both equipment variants.
The following figure will help you to localise the electrical components on the bed.
Product Description
16
Part A: General Information
Содержание Regia
Страница 2: ...Part A General Information ...
Страница 3: ...Regia Standard care bed Regia care bed DSG optional ...
Страница 6: ...II ...
Страница 30: ...Part B Operator and Technical Per sonnel ...
Страница 31: ...Regia Standard care bed Regia care bed DSG optional ...
Страница 36: ...IV ...
Страница 105: ...Appendix Part B Operator and Technical Personnel 69 ...
Страница 106: ...Part C Care staff and residents ...
Страница 107: ...Regia Standard care bed Regia care bed DSG optional ...
Страница 151: ...Cleaning and Disinfection Part C Care staff and residents 41 ...