![burmeier Regia Скачать руководство пользователя страница 88](http://html1.mh-extra.com/html/burmeier/regia/regia_instruction-manual_2832795088.webp)
Inspection report
II. Electrical measurement
(Use only measuring instruments according to DIN EN 62353 (VDE
0751-1))
Note
: To minimise measuring errors, route the test leads as far away as possible from and not par-
allel to the power cables and handset cables of the bed. Also observe the operating instructions for
the measuring instruments used
Insulation resistance
: To be carried out only if there are doubts about the electrical insulation,
such as:
If the customer’s RCD (residual current circuit breaker) has tripped several times
If defective electrical housings are found and at the same time there are signs of spilled liquids/
contamination there that could affect the insulation
1.
Plug the mains cable/switching power supply into the test socket of the measuring instrument
2.
Connect the probe at the common measuring point of all applied parts: = bare screw of the
backrest swivel joint underneath the backrest on the mattress base frame
3.
Start the measuring procedure on the measuring instrument; measuring voltage = 500 V DC
Limit val-
ue
Measured value
Result: Bed prot.
class II (type BF)
≥ 70 MΩ
MΩ
Leakage current (direct or differential current measure-
ment) (type BF)
Proceed as follows:
OK
Not
OK
Description of de-
fect
1.
Plug the mains cable/switch mode power supply into
the test socket of the measuring instrument.
2.
Connect the probe of the measuring instrument to
the bed; measuring point: Bare metal screw under
backrest in frame of mattress base
3.
Operate the motors using the handset for the dura-
tion of the measurement
4.
Start the measurement procedure on the measuring
instrument.
Limit value
Current value
Maintenance
52
Part B: Operator and Technical Personnel
Содержание Regia
Страница 2: ...Part A General Information ...
Страница 3: ...Regia Standard care bed Regia care bed DSG optional ...
Страница 6: ...II ...
Страница 30: ...Part B Operator and Technical Per sonnel ...
Страница 31: ...Regia Standard care bed Regia care bed DSG optional ...
Страница 36: ...IV ...
Страница 105: ...Appendix Part B Operator and Technical Personnel 69 ...
Страница 106: ...Part C Care staff and residents ...
Страница 107: ...Regia Standard care bed Regia care bed DSG optional ...
Страница 151: ...Cleaning and Disinfection Part C Care staff and residents 41 ...