Preparation and setup
Non-compliance with the following rules will result in complete exclusion
of liability on the part of Aesculap.
►
When setting up and operating the product, adhere to:
– national regulations for installation and operation,
– national regulations on fire and explosion protection.
►
Installation and initial commissioning may only be performed by per-
sonnel authorized by Aesculap.
►
Service cover
17
may only be removed by personnel authorized by
Aesculap.
Note
For the safety of patients and users it is essential that the mains power cord
and, especially, the protective earth connection are intact. In many cases
defective or missing protective earth connections are not registered imme-
diately.
►
Observe all specifications regarding ambient conditions, see
Chapter 14.
5.1
Set-up environment/set-up location
The OrthoPilot
®
4 panel PC is approved for use in operating rooms in the
non-sterile area.
The OrthoPilot
®
4 wireless camera is intended to be operated either on the
OrthoPilot
®
4 stand FS110 or on a suitable carrier system.
►
Make sure that the operating elements, power switch and outlet of the
building installation are accessible to the user.
►
Make sure that the plug is accessible to the user.
►
Set up the product in such a way that it can be easily disconnected
from the mains.
►
Lay non-system cables and cords separately from the OrthoPilot
®
4
panel PC. RF cables in particular can cause strong interference.
►
If interference from external devices is suspected, immediately deacti-
vate those devices, if possible, until the source of interference is found.
If necessary, consult a specialist in medical electrical devices to remedy
the interference.
►
If the device is disrupted, shut system down (see Chapter 6.1.2) and
disconnect it from power for approx. 10 seconds.
►
Make certain that the ventilation slots in the back side of the housing
are not covered, e.g. by an OR cloth.
►
Make certain the system is set up on a sufficiently stable support (e.g.
a table, ceiling support, equipment cart, etc.).
►
Follow the instructions for use of the support.
5.2
Mount the OrthoPilot
®
4 panel PC on the
OrthoPilot
®
4 stand/remove the panel PC from the
OrthoPilot
®
4 stand
►
Place OrthoPilot
®
4 stand on a level surface and engage the roller
brakes.
5.2.1
Assembly
►
Insert the OrthoPilot
®
4 panel PC
19
into the self-locking quick-release
holder
20
of the carrier profile
21
, see Fig.
A
.
►
Check interlock: in the locked condition, the locking lever
18
is on the
rear side of the Panel PC in the center position.
►
If desired, secure Panel PC against theft with integrated setscrew on
locking lever
18
.
5.2.2
Disassembly
►
If necessary, loosen the threaded pin on locking lever
18
, see Fig.
A
.
►
Set the locking lever
18
to the right or the left.
►
Grasp the OrthoPilot
®
4 panel PC
19
with both hands and remove ver-
tically upward from the quick-release holder
20
of the carrier
profile
21
.
5.3
Mount the OrthoPilot
®
4 panel PC on a suitable
carrier system/remove from suitable carrier system
5.3.1
Assembly
Disassemble connection interface
The OrthoPilot
®
4 panel PC is mounted on the OrthoPilot® 4 stand with the
connection interface. When using a different carrier system, the connec-
tion interface must be removed.
►
Loosen the four screws
14
and remove together with the VESA quick-
release
15
.
►
Loosen the two screws
16
and remove together with the cover
17
.
►
Loosen the connection cable between the connection interface and
voltage input socket
22
, see Fig.
B
.
Screw the OrthoPilot
®
4 panel PC to the carrier system
►
Screw the OrthoPilot
®
4 panel PC with four M3 screws into the stan-
dard VESA template 75/100 on the carrier system.
►
Attach cover
17
using the two screws
16
.
5.3.2
Disassembly
Remove the OrthoPilot
®
4 panel PC from the carrier system
►
Loosen the four screws in the standard VESA template 75/100 and
remove the OrthoPilot
®
4 panel PC from the carrier system.
Mount connection interface
►
Loosen the connection cable between the connection interface and
voltage input socket
22
, see Fig.
B
.
►
Attach the cover
17
with two screws
16
to the OrthoPilot
®
4 panel PC.
►
Attach the VESA quick release
15
with four screws
14
to the
OrthoPilot
®
4 panel PC.
►
Attach cover
17
using the two screws
16
.
DANGER
Fire and explosion hazard!
►
Use the product outside of explosion hazard
zones (e.g. areas with highly pure oxygen or
anesthetic gases).
DANGER
Risk of death by electric shock!
►
Disconnect power cable before mount-
ing/removal of OrthoPilot
®
4 components.
WARNING
Risk of injury from falling parts!
►
Ensure that the carrier system can securely bear
the weight of the OrthoPilot
®
4 panel PC
(max. 8 kg without connection interface).