“No camera could be detected”
error message
Connection to OrthoPilot
®
4 wireless camera dis-
rupted:
■
Device is switched off.
■
Device is not connected.
Select the “More info” button and follow the instructions for
troubleshooting, see Chapter 7.4.
Connection to OrthoPilot
®
4 wireless camera dis-
rupted:
■
No power supply
■
Internal battery is dead.
Connect to external power supply.
Connection to OrthoPilot
®
4 wireless camera dis-
rupted:
■
Power supply defective.
■
Internal battery defective
■
Camera defective
Have the product repaired by the Aesculap Technical Service.
Connection to OrthoPilot
®
4 wireless camera dis-
rupted:
■
Radio network disrupted
Contact Aesculap Technical Service.
“No foot switch could be detected”
error message
Connection to OrthoPilot
®
4 wireless footswitch
disrupted:
■
Device is switched off
■
Device is not connected
Select the info button and follow the instructions for trouble-
shooting, see Chapter 7.4.
Connection to OrthoPilot
®
4 wireless footswitch
disrupted:
■
Battery dead
■
Issue
■
Footswitch control defective
Finish the navigated operation by data capture via touch-
screen.
Have the product repaired by the Aesculap Technical Service.
“Footswitch seams to be defective”
error message
Connection to OrthoPilot
®
4 wireless footswitch
disrupted:
■
Footswitch software crashed
■
Footswitch control defective
Press the right and left pedals of the foot switch at the same
time for at least 5 seconds to perform a reset and connect to
the footswitch again (select the info button and follow the
instructions for troubleshooting).
If the problem persists, have the product repaired by the
Aesculap Technical Service.
“Battery level is critical” error mes-
sage
Connection to OrthoPilot
®
4 wireless footswitch
disrupted: wireless footswitch battery is nearly
empty.
Have the product repaired by the Aesculap Technical Service.
Error message appears in the selec-
tion of an application tile.
Application does not start, because the desired
application is not registered.
Contact Aesculap Technical Service.
Error message “Patient data not
filled in” appears when selecting an
application tile.
Application is started before patient data was
entered into the OrthoPilot
®
4 home screen.
Enter the patient data, see Chapter 7.1.
Issue
Identification/cause
Remedy