Vercise™ Neural Navigator 4 Programming Manual
Starting a Programming Session
en
4.
Press the
Connect
button next to the Stimulator.
5. Once a connection has been established between the CP and the Stimulator, the following screen will appear (Figure 9 and Figure 10).
Figure 9.
Connection Established Between CP and Stimulator (Vercise Genus or ETS 3)
Figure 10.
Connection Established Between CP and Stimulator (Vercise Gevia, Vercise PC, or ETS 2)
Table 3: Description of the Connect Tab
# Feature
Description
1
Patient ID
Displays the Patient ID number.
2
Connection Status
Displays the connection status between the CP and Stimulator along with the Model and Serial Number of each
of the devices.
#
Connection Status
Description
1
Connection between CP and Vercise Genus or ETS 3 Stimulator
was established.
2
Connection between CP, Programming Wand, and Vercise Gevia,
Vercise PC, or ETS 2 Stimulator was established.
3
Connection not established. Ensure the Vercise Genus or ETS 3
Stimulator is in CP Mode (Pairing Mode). For Vercise Gevia or
Vercise PC, move the Programming Wand closer to the Stimulator
and Scan again, and/or see the
Wand to the CP (Vercise Gevia, Vercise PC, and ETS 2 Stimulators
section of this manual or refer to the
Vercise DBS
Programming Wand Manual
for more information.
3
Refresh Button
Scan for available Stimulators. Disabled if the CP is already connected to the Stimulator.
4
Connection Status
Displays a green or blue filled circle if the Stimulator is connected to the CP.
5
Connect or
Disconnect Button
Connect or disconnect from a Stimulator. When a Stimulator is not connected, this button reads “Connect.”
When a Stimulator is connected, this button reads “Disconnect.”
6 Signal Strength
Displays the signal strength for communication between the CP and Stimulator (Vercise Genus or ETS 3 only).