Programming the Stimulator
en
Viewing the Stimulation Field Model
The Stimulation Field Model (SFM), called STIMVIEW, is a visual representation of the estimated stimulation field for the currently programmed
stimulation parameters. The SFM includes both a visual representation of the DBS Lead as well as the approximated stimulation field shown in
the color red (Figure 14 and Figure 16). As programming parameters are adjusted and the stimulation is steered along the Lead, the SFM will
adjust accordingly.
You may switch between two different views by selecting either 3D Overview or 3D Split View from the Display drop‑down (Figure 16). 3D
Overview presents a three‑dimensional view within which you can zoom, rotate, and pan. The 3D Split View provides a dual pane view centered
on the Lead. The top pane is in‑line with the Lead and the bottom pane is on an axis perpendicular to the Lead. If objects have been imported
from Elements, you may show or hide individual objects using the Object Visibility drop‑down.
Adjust the view of the SFM using
to Zoom,
to Rotate,
to Pan, or
to Reset the original view. When in 3D Split View, both
the Lateral and Axial views of the SFM will adjust in unison using these controls. These controls will not affect or adjust any programming
parameters.
Figure 16.
Display Drop‑down