Estimated Charge Time
The Estimated Charge Time applies to rechargeable Stimulators only. The Estimated Charge Times provides an estimate of the duration and
frequency of charging necessary to maintain stimulation for the selected Program. After the optimal settings have been identified for a Program,
from the
Program Tab
, select
Program Options
and then select
Battery
to obtain the Estimated Charge Time.
Elective Replacement Indicator (ERI) Message
You will not be able to connect to a non‑rechargeable Stimulator that is nearing the end of its battery life. The CP will display the Stimulator
with an ERI message and the Stimulator battery voltage as shown in Figure 27 on the
Connect Tab
. During the ERI period, the Stimulator will
continue to provide stimulation; however, no changes can be made to the Stimulator settings.
Note:
The ERI Message applies to non‑rechargeable Stimulators only.
Note:
The battery voltage shown in Figure 27 is for demonstration purposes only. The ERI indicator varies per Stimulator type.
Figure 27.
ERI Message Displayed on Connect Tab
End of Service (EOS) Message
When the Stimulator has reached its end of service, stimulation will no longer be provided. The CP will display the message shown in Figure 28
on the
Connect Tab
.
Note:
The EOS Message applies to non‑rechargeable Stimulators only.