BT-750 Operator Manual 35
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
●IBP measurement range
Setup Mode
Range of change
Description
WAVE
SPEED
6.5, 12.5, 25,50 mm/sec
Set the speed of IBP waveform
BP1 LABEL
P1 , ABP
Set the level according to the
measurement site of BP. Change the
alarm setting range according to
level.
BP1 SCALE
50,100,200,300 mmHg
Set the display of BP1 wave scale
BP1 ZERO
YES/NO
Adjust the zero point of the
transducer
BP1 ALARM
Refer to Ch.8 Alarm Setup
BP2 LABEL
P2,CVP,PAP,LAP
Set the level according to the
measurement site of BP. Change the
alarm setting range according to
level.
BP2 SCALE
25,50,100,200,300 mmHg
Set the display of BP2 wave scale
BP2 ZERO
YES/NO
Adjust the zero point of the
transducer
BP2 ALARM
Refer to Ch.8 Alarm Setup
IBP
Enable / DISABLE
Select the IBP display
IBP inactive display window (Show the unified NIBP display window.)