BT-750 Operator Manual 14
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
In case of sudden blackout or potable purpose, automatically AC Adaptor power transfers to
battery power for continuous use.
How to use
1.
Battery charge lamp is off, while battery used.
2.
It is durable for approximately 4 hours but depends on battery charged state. ( Full
charged time: 4 hours)
3.
Battery charging starts once main power switch of rear side is on after AC Adaptor
connection.
4.
Battery charged state displays in 5 steps.
Full charged 70% charged 50% charged 20% charged Discharged
5.
When battery is low charged, battery state display on the screen is turned off, beep
sound generates and after about 5 minutes, power is disconnected automatically in
order to protect battery.
Note
When battery electronic pressure is in low state, it may cause
inaccurately measured value. Once battery state is low, connect AC
Adaptor to the appliance at once.
Note
For battery replacement, make sure of polarity.
If the battery polarity is placed incorrectly, it may cause fatal
problem on equipment.
Contact to Bistos head quarter or authorized agent, when you need
battery replacement.
If a side of power cord is put to power port of BT-750, power is supplied to the equipment. Make sure
that each cord is connected to input / output terminal correctly. If there is incorrect connection,
it may cause problem.
※
Battery Charge
If AC power is disconnected, power is supplied thru battery. When battery power is not
sufficient, you should charge it by connecting AC adaptor.
Recharging time : More than 5 hours.
Duration time: Under 3 hours.
In the case of discharge battery.
Caution
Noise may occur in waveform, while using AC power. In this case,
ground the product.
It may occur the fatal damage to the product ,while connecting AC
power with wet hand.
Caution
While the monitor is working, battery charging lamp is on. When
sub-power switch is turned off and battery charging completes, lamp
is out.
※
Battery type
BT-750 uses Lithium-lon battery.