BT-750 Operator Manual 29
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
5.4 ECG1, ECG2 Setup
5.4.1 ECG Menu Display
1) “Press the button “
” to go to adjustment display
2) Select the menu to change with trim knob and press the trim knob to set the location to
be modified. Then, selected menu is displayed.
3) Turn to right and left with trim knob and press the trim knob to set up the numeric value
and save it.
4) Press “Return” button to go back to 1
st
display window.
5) Press the button “
” to return to main display.
Setup Mode
Range of modification
Description
ECG 3 LEAD
LEAD I, II, III
Set ECG Lead
ECG 5 LEAD
LEAD I, II, III, aVR, aVL,
aVF, V(Chest)
ECG GAIN
1/2,1,2 times
Set the size of ECG wave
WAVE SPEED
6.5, 12.5, 25, 50 mm/sec
Set the velocity of ECG wave
ECG MODE
Monitor/Operation/Diagnosis
Set Analog and Digital filter in ECG
wave
LEAD MODE
3 LEAD, 5 LEAD
Set ECG Cable
ARRHYTHMIA
ON/OFF
Set the usage of Arrhythmia analysis
PACEMAKER
ON/OFF
Set the detection of the pacemaker
PVC ALARM
ON/1~30
Select whether the alarm generates
or no on detection of PVC .
HR ALARM
Refer to Ch.8 Alarm setup
ST
SEGMENT
P-R /J + 4~252 ms ,Alarm
ON/OFF
Set P-R, J Point