BT-750 Operator Manual 31
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
5.4.2.1 ST Analysis
(When ST Segment is on)
ST Level is measured by horizontal comparison between PR-segment and ST-segment.
User can adjust J point and P-R point at 1~250 4m/sec interval. These measured values
are available only in normal heart rate. In case of abnormal and pacemaker, for example,
Premature Ventricular Contraction, Ventricular Tachycardia etc., it is except ST analysis
because it is possible to cause distortion of ST form.
1) “Press the button “
” to go to adjustment display
2) Select the menu to change with trim knob and press the trim knob to set the location to
be modified. Then, selected menu is displayed.
3) Turn to right and left with trim knob and press the trim knob to set up the numeric value
and save it.
4) Press “Return” button to go back to 1
st
display window.
5) Press the button “
” to return to main display.
● Arrhythmia Analysis can judge problem of Ventricle.
Caution
Our ST measurement is only on the basis of ECG1.
5.4.2.2 PACEMAKER Detection
(When PACEMAKER is on)
● In case Pacemaker is Detection ON, red bar indicate it on the ECG wave form periodically
from input of pacemaker detection