BT-750 Operator Manual 42
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Caution
Inspection does not carry out the patients with arrhythmia, premature
ventricular contraction or ventricular tachycardia.
6.2 Respiration Menu Display
1) Press the button “
” to go to adjustment display.
2) Select the menu to change with trim knob and press the trim knob to set the location to
be modified. Then, selected menu is displayed.
3) Turn to right and left with trim knob and press the trim knob to set up the numeric
value and save it.
4) Press “Return” button to go back to 1
st
display window.
5) Press the button “
” to return to main display.
The range of respiration setup
Setup Mode
Range of modification
Description
WAVE SPEED
6.5/12.5/25 mm/s
Set Wave Speed
RESP LEAD
RA-LL/RA-LA
Set one of two
RESP GAIN
ⅹ
1/2,
ⅹ
1,
ⅹ
2
Set the size of RESP waveform
APNEA ALARM
ON/OFF
Set alarm for apnea
APNEA TIME [SEC]
20~120 Sec.
Set detection of Apnea
RESP ALARM
Refer to CH.8 Alarm Setup