BT-750 Operator Manual 58
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
NIBP Module
Cuff Inflation Rate
Auto mode Repetition
Measurement Time (Typical at HR > 60bpm)
Accuracy
Maximum Standard Deviation
Maximum Mean Error
Adult / pediatric Mode
Measurement Ranges
Systolic
Diastolic
MAP
Neonate Mode
Measurement Ranges
Systolic
Diastolic
MAP
Overpressure Alarm Limit value
Adult
Neonate
3 to 18 seconds
1,3,5,10,30,60,90,120,240 minutes
25 seconds
±8 mmHg
±5 mmHg
50 ~ 255 mmHg
30 ~ 220 mmHg
40 ~ 235 mmHg
30 ~130 mmHg
20 ~ 100 mmHg
25 ~ 120 mmHg
300 mmHg(40kPa)
150 mmHg(20kPa)
Respiration rate
Method Type
Range
Lead
Accuracy
Display Sweep Speed
Lead off Condition
Trans-thoracic impedance
0 to 150 breaths/min
Lead 2 only RA to LA
±1 breaths/min
6.25 mm/s, 12.5 mm/s, 25.0 mm/s
Detected and displayed
Temperature
Method Type
Compatible with
Parameter displayed
Range
Display Accuracy
Thermistor
YSI Series 400 Temperature Probes
TEMP1, TEMP2
10
℃
to 45
℃
[50
℉
to 113
℉
]
±0.1
℃
(25
℃
to 45
℃
) or ±0.2
℉
[77
℉
to
113
℉
]
±0.2
℃
(15
℃
to less than 45
℃
)[to
℉
] or ±0.4
℉
[59
℉
to less than 77
℉
]