BT-750 Operator Manual 37
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
5.9 CO2 Setup
1) Press the button “
” to go to adjustment display.
2) Select the menu to change with trim knob and press the trim knob to set the location to
be modified. Then, selected menu is displayed.
3) Turn to right and left with trim knob and press the trim knob to set up the numeric
value and save it.
4) Press “Return” button to go back to 1
st
display window.
5) Press the button “
” to return to main display.
● Setup range of EtCO2
Setup Mode
Range of Change
Description
WAVE
SPEED
6.5, 12.5, 25 mm/sec
Set the speed of EtCO2 waveform
UNIT
mmHg, kPa, %
Set the unit of EtCO2
measurement
SCALE
~12,~25, ~50, ~70, ~100,~150
mmHg
Set the WAVE SCALE
ZERO
YES / NO
Adjust the zero point of the
transducer
AVERAGE
1 BREATH, 10, 20 Sec
Set DATA average time DATA
Gas COMP
NON,O2/N2O, N2O, O2
Set the Gas type
EtCO2
Enable / DISABLE
Set ON/OFF of EtCO2 module
ALARM
Refer to Ch.8 Alarm Setup