BM3 User’s Manual
WARNING
In the monitoring of patients the coincidence of adverse conditions may lead to a disturbed
signal going unnoticed. In this situation artifacts are capable of simulating a plausible
parameter reading, so that the monitor fails to sound an alarm. In order to ensure reliable
patient monitoring, the proper application of the probe and the signal quality must be
checked at regular intervals.
SPO2 Settings
A.
SPO2 menu
MENU
Description
Available Settings
A-1.
Alarm
SPO2 Alarm setup menu
A-1-1.
PARAMETER ALARM
LIMIT
PERCENT, PR parameter alarm , level ,
activate setup menu
A-1-2.
TECHNICAL ALARM
CONDITION
SPO2-PROBEOFF
SPO2-CHECKPROBE
SPO2-POORSIGNAL
SPO2-LOSTPULSE
SPO2-ARTIFACT
SPO2-PULSE SEARCH
A-2.
RATE VOLUME
Menu in which RATE VOLUME is set up
When the ECG volume is set, it is
automatically set to OFF.
OFF, 0%~100%
A-3.
DISPLAY OPTION
SPO2 waveform display setting
A-3-1.
SWEEP SPEED
It can set the speed of SPO2 displayed
on the screen. Default: 25 mm/s.
6.25mm/s, 12.5mm/s,
25mm/s, 50mm/s