Rev. 3.0
16
BM3 User’s Manual
Caution
The Equipment should be placed far from generator, X-ray equipment, broadcasting
equipment or transmitting wires, so as to prevent the electrical noises from being
generated during the operation, When these devices are near the Equipment, it can
produce inaccurate measurements. For BM3 both independent circuit and stable
grounding are essentially required. In the event that the same power source is
shared with other electronic equipment, it can also produce inaccurate output.
Note
BM3 is classified as follows:
- BM3 classifies as Class
II,
BF
&
CF concerning electric shock. It is not proper to
operate this Equipment around combustible anesthetic or dissolvent.
- Noise level is A class regarding IEC/EN 60601-1 and the subject of Nose is A level
concerning IEC/EN60601-1-2.
Warning
Do not touch the patient while using the defibrillator. The user may be at risk.
When using the defibrillator, be careful about safety and use only the supplied
cable.
Warning
In case the Equipment does not operate as usual or damaged, do not use on
patient, and contact to the medical equipment technician of the hospital or the
equipment supply division.