Guidance and manufacturer’s declaration - electromagnetic immunity
The
BM3
system is intended for use in the electromagnetic environment specified below.
The customer or the user of the
BM3
system should assure that it is used in such an environment
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic environment -guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80MH
z
3 Vrms
150 kHz to 80 MH
z
BM3
system must be used only in a shiel
ded location with a minimum RF shielding
effectiveness and, for each cable that ent
ers the shielded location with a minimum
RF shielding effectiveness and, for each c
able that enters the shielded location
Radiated RF
IEC 61000-4-3
3 V/m
80.0 MHz to 2.5
GHz
3 V/m
80.0 MHz to 2.5 G
Hz
Field strengths outside the shielded locati
on from fixed RF transmitters, as determin
ed by an electromagnetic site survey, sho
uld be less than 3V/m.
a
Interference may occur in the vicinity of e
quipment marked with the following sym
bol:
Note 1)
These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Note 2)
It is essential that the actual shielding effectiveness and filter attenuation of the shielded
location be verified to assure that they meet the minimum specification.