Rev. 3.0
117
BM3 User’s Manual
the measurement.
the type of gas that is mixed with the
breathing gas measuring
N2O
HELIUM
A-5-3.
SLEEP MODE
Sleep mode is used to save power
when the host monitor is in standby
mode. There are two sleep modes
available for the Capnostat. Using
Sleep Mode 1 maintains the heaters so
the Capnostat is able to run
immediately after exiting the sleep
mode. Mode 2 will require the
Capnostat to go through its warm up
sequence when exiting this mode and
a delay will be introduced until the
system has stabilized.
NORMAL MODE
TURNOFF MODE
POWER SAVING
A-5-4.
BARO. PRESSURE
This setting is used to set current
Barometric Pressure.
760mmHg
A-5-5.
GAS TEMPERATURE
This setting is used to set temperature
of the gas mixture. This setting is
useful when bench testing using static
gasses where the temperature is often
room temperature or below.
35.0 °C
A-5-6.
O2 COMPENSATION
Use this setting to correct for the
compensation of the gas mixture
administered to the patient.
A-5-7.
ANESTHETIC AGENT
Anesthetic agent is ignored when the
balance gas is set to helium.
A-5-8.
ZERO TYPE
This function is used to initiate a
When performing a zero on room air,
this setting should be set to room air
ROOM AIR
N2