Rev. 3.0
101
BM3 User’s Manual
turned on and pressurized to the
maximum pressure range at the initial
pressurization.
Default Settings value:
ADT : 170 mmHg
PED : 140mmHg
NEO : 120mmHg
80 – 170mmHg
NEO :
60 – 140mmHg
A-4.
SETTING TIME
How to apply pressure value setting.
Once: When the blood pressure is
measured for the first time, the
pressure is set to the set pressure
value, but automatically adjusted
according to the patient's blood
pressure value.
Every Time: Whenever blood pressure
is measured, pressurize to the set
pressure value every time
Once,
Every Time
A-5.
AUTO
MEASUREMENT
INTERVAL
A menu to set Interval time when
measures the blood pressure
periodically.
After setting INTERVAL, you must press
NIBP KEY to start NIBP START
periodically.
1min, 2, 3, 4, 5, 10,
15, 20, 30, 1hour, 2,
4, 8
B-1.
NIBP STAT
Patients with severe state changes in
blood pressure are in continuous
mode for 5minutes to check for
changes in blood pressure
continuously.
OFF / ON
C-1.
VITAL SIGN REVIEW
Record the 40 most recently measured
blood pressure values.