Rev. 3.0
133
BM3 User’s Manual
Noise in ECG
-
Gel is dry
-
Electrodes does not stick well to skin
The patient's skin
is extremely dry
Apply ECG gel or water to the area of
contact then use the electrode
Relpace the electrodes
(Use CE compatable products)
Replace lead wires
Repair the ECG B/D
No
Yes