Note
In the following monitoring conditions, the measured values may be inaccurate. Read the
measured values carefully.
1. When using this in an environment of using nitrous oxide gas of high concentration
2. When using this in an environment where abrupt temperature change takes place
3. When using this in an environment with severely high humidity.
Caution
The measured values may be inaccurate when using this equipment for patients who
have very fast or irregular respiration.
When measuring CO2 from the patient under the anesthesia, check it when gas mixture
comes in. Otherwise, the measured result values may be inaccurate.
When using a anesthesia machine that uses a volatile anesthetic, CO2 values may be
inaccurate.