Rev. 3.0
27
BM3 User’s Manual
Warning
Older or defective batteries will have significantly reduced capacity or operating
time.
note
To maximize the charge for transport, keep the monitor connected until
you are ready to transport the patient. Reconnect the monitor immediately
after transport.
Bionet recommends replacing the lithium ion battery after 24 months of
use.
Battery life depends on usage. If battery life continues, battery life will
decrease and frequency of replacement will increase.
To prevent pre-discharge, recharge after the battery is discharged.
Caution
The battery charge display is accurate only when the battery is operating normally.
- Battery Charging Time: more than 6 hours
- Continuous Battery Usage Time: more than 3 hours for normal operation.(when fully
charged)
Caution
The battery charge display is displayed correctly only when the battery is operating
normally
Note
If no AC power is applied, the battery charge display will take up to 15 seconds to
reflect the actual capacity of the internal battery.