Rev. 3.0
81
BM3 User’s Manual
Arrhythmia Settings
A.
ECG menu
MENU
Description
Available Settings
A-1.
Arrhythmia
ARRHYTHMIA parameter alarm,
level, Activation setup menu.
A-1-1.
Arrhythmia Type
Sets up ON/OFF to indicate
detection of diagnosis (Asys,
VTAC/VFIB and VTAC).
OFF: Do not perform arrhythmia
diagnosis.
LETHAL: Performs the detection of
Asys, VTAC/VFIB, and VTAC at the
selected lead
OFF, LETHAL
A-1-2.
Arrhythmia Alarm
Condition
Alarm setting menu by arrhythmia
type
Warning
Display Heart Beat Equipment Signal
Hart Beat equipment signal displays when the PACE mode is. the signal appears
series form. The signal size or form are meaningless clinically
Number Of Heart Beat
Attention to the patient with heart beat equipment. The heart beat equipment can
show heart beat even during arrhythmia continuously. Therefore, do not depend on
heart beat alarm excessively.