Chapter 7 Creating Experiments and Workspaces
n
After sample boost, the software runs an additional 5 µl of sample to stabilize flow before
acquisition begins.
n
The probe remains in the sample while acquiring the target volume.
n
Return Sample is automatically activated. Excess sample in the line returns to the tube or well
after target volume is acquired.
Important:
Users can set Return Sample to OFF if necessary. However, when using
Volumetric Counting with Return Sample set to OFF, users will experience sample loss. If
sample loss is a concern, run Volumetric Counting with Return Sample set to ON.
Return Sample
When Return Sample is activated, any sample remaining in the sample line after the required number
of events has been acquired is returned to the tube/well.
Flow Rate or Event Rate Settings
The flow rate controls allow you to specify a precise target flow rate for each position, in µl/sec.
The following flow rate options are available:
n
Low
— 0.5 µl/sec (30 µl/min)
n
Normal
— (default) 1.0 µl/sec (60 µl/min)
n
High
— 1.5 µl/sec (90 µl/min)
Note:
The flow rates values are not editable.
Alternatively, you can specify a target event rate, in events/sec (e/s).
During acquisition in standard mode, you can adjust the target flow rate manually (on the instrument)
from 0.1 µl/sec up to 3.5 µl/sec. However, you cannot adjust the flow rate during acquisition in high-
throughput mode.
Note:
In high-throughput mode, all samples must use the same target flow rate.
196 | ZE5 Cell Analyzer and Everest Software
Содержание ZE5
Страница 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Страница 2: ......
Страница 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Страница 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Страница 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Страница 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Страница 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Страница 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Страница 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Страница 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Страница 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Страница 125: ...Working with Optical Filter Configurations User Guide 123 ...
Страница 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Страница 173: ...Setting Up the Run List The plate setup displays a layout of the selected media User Guide 171 ...
Страница 177: ...Setting Up the Run List User Guide 175 ...
Страница 191: ...Setting Up the Run List User Guide 189 ...
Страница 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Страница 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Страница 333: ...Decontaminating the System User Guide 331 ...
Страница 380: ...Appendix B Example 9 Color Immunophenotyping Experiment 378 ZE5 Cell Analyzer and Everest Software ...
Страница 384: ...Appendix C Optional External DI Water and Waste Ports 382 ZE5 Cell Analyzer and Everest Software ...
Страница 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Страница 397: ......